Calibration schedules & audit readiness
Reference guides for biomedical departments, administrators and quality managers.
RESOURCES
Standard Calibration Frequency Guide
General recommendations by clinical risk category.
| Equipment | Risk class | Recommended cycle | Critical parameters |
|---|---|---|---|
| ICU Ventilator | High (Life Support) | Every 6 Months | Tidal volume, PEEP, FiO₂, flow sensor |
| Defibrillator | High (Emergency) | Every 6 Months | Delivered energy (J), sync time |
| Infusion / Syringe Pump | Medium-High | Annual (6–12 Mo) | Flow accuracy, occlusion pressure |
| Anaesthesia Machine | High (Surgical) | Every 6 Months | Vaporizer output, hypoxic guard, pressure |
| Patient Monitor (ECG/SpO₂) | Medium | Every 12 Months | Lead rhythm, SpO₂ curve, NIBP |
| Infant Radiant Warmer | High (Neonatal) | Every 6 Months | Probe accuracy, irradiance, alarms |
General recommendations; follow manufacturer and regulatory requirements.
AUDIT READINESS
Biomedical Audit Checklist
Asset Register
Inventory with unique identification numbers.
Valid Calibration Certificates
Traceability and next-due date.
Electrical Safety Reports
IEC 60601 earth leakage and ground continuity logs.
Preventive Maintenance Logs
Scheduled service records signed by qualified personnel.